# VIDEO APPLICATION

# GC/MS/MS Analysis of PAHs Using Hydrogen Carrier Gas with the Agilent HydroInert Source in a Challenging Soil Matrix

![first_slide](https://webinar.sepscience.com/hs-fs/hubfs/SS_VideoApp_Agilent_IO__Jun2023/first_slide.png?width=360&height=203&name=first_slide.png)

Separation Science, in collaboration with   
Agilent, offers a video application and complementary application notes focused on GC/MS/MS analysis of PAHs.

**Format:** Video Application

**Duration:**  13 minutes

Presenter:  
![SamuelHaddad_200](https://webinar.sepscience.com/hs-fs/hubfs/Webinars/Agilent_220223/SamuelHaddad_200.png?width=80&height=80&name=SamuelHaddad_200.png)Samuel Haddad, PhD  
*Application Development Chemist, Agilent Technologies *  
Samuel Haddad is a GC Application Development Chemist at Agilent Technologies in Little Falls, Delaware. Samuel specializes in GC and GCMS applications of trace analysis in environmental and heavy matrix samples. His specialty includes sample preparation techniques to separate analytes from matrix components and extraction from a wide variety of matrices including water, fish, aquatic life, soil, and complex fluids. Samuel holds a Ph.D. in Environmental Science from Baylor University and a B.S. in Biology from Southern Illinois University Edwardsville.

**Sponsor:**

  
**![Agilent_Logo_RGB](https://webinar.sepscience.com/hs-fs/hubfs/Companies/Agilent/Agilent_eBook_July2021/Agilent_Logo_RGB.png?width=200&name=Agilent_Logo_RGB.png)**

 

[![Share on Facebook](https://static.hubspot.com/final/img/common/icons/social/facebook-24x24.png)](http://www.facebook.com/share.php?u=https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Dfacebook) [![Share on LinkedIn](https://static.hubspot.com/final/img/common/icons/social/linkedin-24x24.png)](http://www.linkedin.com/shareArticle?mini=true&url=https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Dlinkedin) [![Share on Twitter](https://static.hubspot.com/final/img/common/icons/social/twitter-24x24.png)](https://twitter.com/intent/tweet?original_referer=https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Dtwitter&url=https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Dtwitter&source=tweetbutton&text=GC%2FMS%2FMS%20Analysis%20of%20PAHs%20Using%20Hydrogen%20Carrier%20Gas) [![Share on Email](https://static.hubspot.com/final/img/common/icons/social/email-24x24.png)](mailto:?subject=Check%20out%20https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Demail%20&body=Check%20out%20https%3A%2F%2Fwebinar.sepscience.com%2Fform%2Fview%2Fgc-ms-ms-analysis-of-pahs-using-hydrogen-carrier-gas-with-the-agilent-hydroinert-source-in-a-challenging-soil-matrix%3Futm_medium%3Dsocial%26utm_source%3Demail)

## Presentation Overview

Whilst helium is the gold standard that generally offers the best performance for GC/MS analysis, with proper configuration and conditions, comparable results can be achieved with hydrogen. PAH analysis is a great application to consider for hydrogen as the performance not only equals helium, but for some compounds can even exceed helium results due to chromatographic improvements seen with hydrogen. For laboratories looking to shift their PAH analysis to the more sustainable hydrogen carrier gas, the HydroInert source with a 9 mm extractor lens enables the transition with equivalent or better performance for GC/MS and GC/MS/MS analysis.

This presentation describes GC/MS/MS analysis of PAHs using hydrogen carrier gas with the Agilent HydroInert source in a challenging soil matrix. The Agilent 8890/7000E triple quad and 8890/5977C single quad GC/MS systems with a Hydroinert source, optimized for hydrogen carrier gas, were used for the analysis of polycyclic aromatic hydrocarbons (PAHs).

The optimized method using the HydroInert source provides excellent peak shape, especially for the late-eluting PAHs. Calibration linearity of R2 ≥ 0.999 was observed for all 27 analytes over <1 – 1,000 pg. Method detection limits were 0.1 pg on average. Internal standard (ISTD) response was stable across 4 orders of calibration with RSDs below 7%. Stable performance over 500 injections of a challenging soil extract demonstrated quantitation stability with RSDs <12% for all analytes with routine maintenance.

From this presentation and associated application notes, you will learn about:

- the novel HydroInert source
- considerations when switching to hydrogen carrier gas
- a novel application for the analysis of PAHs using hydrogen carrier

In addition to the video application, you will be able to download your personal copy of the associated application notes and user guide.

![app_note_gcmsms_thumb](https://webinar.sepscience.com/hs-fs/hubfs/SS_VideoApp_Agilent_IO__Jun2023/app_note_gcmsms_thumb.png?width=80&height=100&name=app_note_gcmsms_thumb.png)

GC/MS/MS Analysis of PAHs with Hydrogen Carrier Gas

![app_note_pah_analysis_thumb](https://webinar.sepscience.com/hs-fs/hubfs/SS_VideoApp_Agilent_IO__Jun2023/app_note_pah_analysis_thumb.png?width=80&height=100&name=app_note_pah_analysis_thumb.png)

Analysis of PAHs Using GC/MS with Hydrogen Carrier Gas and the Agilent HydroInert Source

![app_note_eigcms_thumb](https://webinar.sepscience.com/hs-fs/hubfs/app_note_eigcms_thumb.png?width=80&height=100&name=app_note_eigcms_thumb.png)

Agilent EI GC/MS Instrument Helium to Hydrogen Carrier Gas Conversion - User Guide

 

To register to view this video application and download the associated application note and user guide PDFs, please complete the form on the right. 

Registrants to this video application are covered by the Eclipse Business Media [Privacy Policy](http://www.sepscience.com/privacy-policy). Information you provide will be shared with the sponsors for this content. Eclipse Business Media Ltd may contact you to offer you content based on your specific interests. You may opt-out at any time.

Eclipse Business Media Ltd, Regd in England, No. 06513189, Woodview, Bull Lane Industrial Estate, Sudbury, CO10 0FD, United Kingdom, T +44 (0)161 818 7434, [info@sepscience.com](mailto:info@sepscience.com)

 

Eclipse Business Media Ltd needs the contact information you provide to us to contact you about our products and services. You may unsubscribe from these communications at any time. For information on how to unsubscribe, as well as our privacy practices and commitment to protecting your privacy, check out our [Privacy Policy](https://www.sepscience.com/privacy-policy/) or contact [privacy@labxmediagroup.com](mailto:privacy@labxmediagroup.com) 

Copyright © 2025. All rights reserved

[Contact Us](https://www.sepscience.com/contact-us/)  
[Advertise with Us](https://www.sepscience.com/advertise-with-us/)